Z-0198-2007
StatSpin Express 3 Centrifuge with RTX8 Rotor Model: M502
StatSpin, Inc., dba Iris Sample Processing / initiated 2006-09-29 / Terminated
Iris Sample Processing began a recall of StatSpin Express 3 Centrifuge with RTX8 Rotor Model: M502 on . The reason FDA records is "Rotor may crack and separate causing device to fail. A failed rotor not successfully contained may seriously injure the laboratory worker". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0198-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- StatSpin, Inc., dba Iris Sample Processing
- Product
- StatSpin Express 3 Centrifuge with RTX8 Rotor Model: M502
- Product code
- JQC Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use
- Reason for recall
- Rotor may crack and separate causing device to fail. A failed rotor not successfully contained may seriously injure the laboratory worker.
- FDA root cause
- Other
- Action
- Iris telephoned customers on 9/29/06 and advised them of the recall of the rotor. New rotors were mailed Overnight Confirmed Delivery with instructions for replacement and return of the the old rotor.
- Quantity in commerce
- 94 units (Revised on 11/15/06 to 116 units)
- Distribution
- Nationwide
- Product codes and lots
- Rotors with serial numbers: B1562 to B1738