M3001A Philips Multi Measurement Server (MMS)
Philips North America LLC / initiated 2006-09-21 / Terminated
Philips Medical Systems began a recall of M3001A Philips Multi Measurement Server (MMS) on . The reason FDA records is "Unexpected pulse oximetry (Sp02) readings (100%) over time when a sensor is not attached to the patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0183-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- M3001A Philips Multi Measurement Server (MMS)
- Product code
- DQA Oximeter
- Reason for recall
- Unexpected pulse oximetry (Sp02) readings (100%) over time when a sensor is not attached to the patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM
- FDA root cause
- Other
- Action
- Philips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory.
- Quantity in commerce
- 69696 units
- Distribution
- Worldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Germany, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Mauritania, Mauritius, Netherlands, Norway, Oman, Slovenia, South Africa, Spain, Switzerland, Turkey, UAE, UK, Yemen, Zambia, Zimbabwe
- Product codes and lots
- HW Revision B: M3001 A #AO1 with S/N DE512xxxx M3001A #A02 with S/N DE512xxxx HW Revision C M3001A #A01 and #A02 with S/N DE610xxxx With Sp)2 Firmware (Fw) Rev.: Sp02 DSP FW-Rev: A.01.41 or A.01.42