Z-0433-2007

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required)...

Terumo Cardiovascular Systems Corporation / initiated 2006-10-16 / Terminated

Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the... on . The reason FDA records is "A 'battery needs service' message may be displayed, or the remaining battery time may incorrectly become zero, even when the battery is fully functional". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0433-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
A 'battery needs service' message may be displayed, or the remaining battery time may incorrectly become zero, even when the battery is fully functional.
FDA root cause
Other
Action
Consignees were sent a letter dated 11/10/06 informing them of the nature of the issues and that the firm would contact them to schedule a service call to install the software upgrade package. The software upgrade installation began at customers on 10/16/06, prior to issuance of the letter to consignees.
Quantity in commerce
207
Distribution
Worldwide distribution ---- including USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Product codes and lots
Serial numbers 0006 through 0066 and 0100 through 0245.