Z-0186-2007

ambIT Ambulatory Infusion Pump Cassette; Filter, Spike, Sterile, Vol. 2.3ml, REF 220140, Sorenson Medical...

Sorenson Medical, Inc. / initiated 2006-10-10 / Terminated / root cause: sterility

Sorenson Medical, Inc. began a recall of ambIT Ambulatory Infusion Pump Cassette; Filter, Spike, Sterile, Vol. 2.3ml, REF 220140, Sorenson Medical, Inc., West Jordan, Utah on . The reason FDA records is "Sterile disposable cassette used with infusion pump may cause an under-infusion when used in a low flow rate therapy of less than 10 ml per hour". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0186-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Sorenson Medical, Inc.
Product
ambIT Ambulatory Infusion Pump Cassette; Filter, Spike, Sterile, Vol. 2.3ml, REF 220140, Sorenson Medical, Inc., West Jordan, Utah.
Product code
FRN Pump, Infusion
Reason for recall
Sterile disposable cassette used with infusion pump may cause an under-infusion when used in a low flow rate therapy of less than 10 ml per hour.
Root cause tag
sterility
FDA root cause
Other
Action
Consignees were notified by letter on 10/10-11/2006 and told to immediately discontinue use/quarantine all affected cassettes. .
Quantity in commerce
30,697 units
Distribution
Worldwide, including USA, Belgium, France, Germany, Italy, Singapore, Venezuela.
Product codes and lots
REF 220140, all serial numbers