Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0082-2007

TOSOH AIA-PACK FOLATE Calibrator Sets, six levels, Catalog Number-020392. The calibrators are packaged in 1...

Tosoh Bioscience Inc / initiated 2006-09-01 / Terminated / root cause: manufacturing process

Tosoh Bioscience Inc began a recall of TOSOH AIA-PACK FOLATE Calibrator Sets, six levels, Catalog Number-020392. The calibrators are packaged in 1 ml glass vials with 12 glass vials per box on . The reason FDA records is "Reports of low calibration rate value flags; Use of the recalled product may result in failure to calibrate the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0082-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tosoh Bioscience Inc
Product
TOSOH AIA-PACK FOLATE Calibrator Sets, six levels, Catalog Number-020392. The calibrators are packaged in 1 ml glass vials with 12 glass vials per box.
Product code
JIS Calibrator, Primary
Reason for recall
Reports of low calibration rate value flags; Use of the recalled product may result in failure to calibrate the instrument
Root cause tag
manufacturing process
FDA root cause
Other
Action
The firm mailed Urgent: Medical Device Recall letters to their customer accounts on 9/5/2006. The letter informs the customers of the calibration problem and asks that the customers examine their inventories for the presence of the suspect product and place any product found in quarantine. The letter also asks that the customers identify their subaccount/customers and notify them of the recall with a copy of the recall letter, if the product has been further distributed. The recall letter is accompanied by a Response form which the customer is asked to complete and return to TOSOH as soon as possible. The Response form requires reporting of infomation concerning the amount and disposition of all product on-hand. It also includes a section for use by the distributor/customers for indicating whether or not their customers have been notified as well as the names and addressed of the distributor''s subaccounts. Should the customer(s) have any questions, the recall letter provides a telephone number and the name of a contact person to whom questions can be addressed.
Quantity in commerce
151 boxes
Distribution
Nationwide distribution ---- including states of AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, OR, PA, PR, SC, TX, WA, WI, WV, and WY.
Product codes and lots
Lot Code: G431726 (shipped after June 6, 2006)