Z-0160-2007
Fusion OMNI ERCP Catheter, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC...
Wilson-Cook Medical Inc. / initiated 2006-10-05 / Terminated / root cause: manufacturing process
Cook Endoscopy began a recall of Fusion OMNI ERCP Catheter, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105 on . The reason FDA records is "Injection through the flush port of these ERCP catheters may be compromised due to omission of a manufacturing activity". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0160-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Wilson-Cook Medical Inc.
- Product
- Fusion OMNI ERCP Catheter, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105
- Product code
- KOG
- Reason for recall
- Injection through the flush port of these ERCP catheters may be compromised due to omission of a manufacturing activity.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Cook Ireland customers were notified by letter via email on 10/13/2006.
- Quantity in commerce
- 20 units
- Distribution
- Worldwide.
- Product codes and lots
- Lot number W2243511, Expiration date 7/12/2009
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.