Invasive Blood Pressure Interface Cables: a) Medrad 9500 Multigas monitor for MRI, Invasive Blood Pressure...
Medrad Inc began a recall of Invasive Blood Pressure Interface Cables: a) Medrad 9500 Multigas monitor for MRI, Invasive Blood Pressure Interface Cable (Part # 3006314) for use with... on . The reason FDA records is "Interference problems: certain invasive blood pressure transducers connected to these cables are susceptible to interference if exposed to static magnetic field strengths exceeding 300 gauss. In certain orientations...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0258-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medrad, Inc.
- Product
- Invasive Blood Pressure Interface Cables: a) Medrad 9500 Multigas monitor for MRI, Invasive Blood Pressure Interface Cable (Part # 3006314) for use with Abbott/Hospira IBP Transducer b) Invasive Blood Pressure Interface Cable (Part # 3008772) for use with Braun IBP Transducers. The product is an accessory to patient physiological monitor.
- Product code
- MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
- Reason for recall
- Interference problems: certain invasive blood pressure transducers connected to these cables are susceptible to interference if exposed to static magnetic field strengths exceeding 300 gauss. In certain orientations, the transducer causes the invasive blood pressure readings on the physiological monitor to display inaccurate readings.
- FDA root cause
- Other
- Action
- The recalling firm contacted all affected customers via telephone or letter dated November 20, 2006 to inform them of the problem and instruct them to return the cables or apply corrective labeling.
- Quantity in commerce
- 80 units
- Distribution
- Worldwide distribution ---- including USA state of CA, CO, HI, IL, IN, MA, MD, NJ, NY, OH, PA, TX, and WA and countries of The Netherlands, Denmark, Australia, Brazil, Italy, and Japan.
- Product codes and lots
- Catalog number 3006314 and 3008772