FastCath Transseptal Catheter Introducers with Hemostasis Valve, DAIG division of St. Jude Medical...
St. Jude Medical / Daig Division began a recall of FastCath Transseptal Catheter Introducers with Hemostasis Valve, DAIG division of St. Jude Medical, Minnetonka, MN, consists of a radiopaque sheath and... on . The reason FDA records is "Limited number of FastCath Transseptal Catheter Introducers (REF #406850) are susceptible to the irrigation extension tube on the side port becoming disconnected when a pull force is applied". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0535-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- FastCath Transseptal Catheter Introducers with Hemostasis Valve, DAIG division of St. Jude Medical, Minnetonka, MN, consists of a radiopaque sheath and dilator; Product # 406850.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Limited number of FastCath Transseptal Catheter Introducers (REF #406850) are susceptible to the irrigation extension tube on the side port becoming disconnected when a pull force is applied.
- FDA root cause
- Other
- Action
- The affected customers were notified on 10/5/06 by letter.
- Quantity in commerce
- 40 Units
- Distribution
- Nationwide.
- Product codes and lots
- Lot number 1239904