Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0248-2007

Sling Rope/Connector Assembly Kits for the Sara and Sara Nova standing and raising aids; the kit is comprised...

Arjo, Inc. / initiated 2006-10-30 / Terminated / root cause: manufacturing process

Arjo, Inc. began a recall of Sling Rope/Connector Assembly Kits for the Sara and Sara Nova standing and raising aids; the kit is comprised of 2 x 200 cm ropes, 2 knobs, 2 sling connectors... on . The reason FDA records is "The sling ropes were manufactured with insufficient rope length past the knot to prevent the knot from coming loose". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0248-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arjo, Inc.
Product
Sling Rope/Connector Assembly Kits for the Sara and Sara Nova standing and raising aids; the kit is comprised of 2 x 200 cm ropes, 2 knobs, 2 sling connectors, 2 socket cap screws, 2 nylon nuts and 2 spacers; Arjo Inc., 50 N. Gary Avenue, Roselle, IL 60172; Model KS1006
Product code
FSA Lift, Patient, Non-Ac-Powered
Reason for recall
The sling ropes were manufactured with insufficient rope length past the knot to prevent the knot from coming loose.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Arjo Inc. USA sent recall letters to the direct accounts on 10/30/06, informing them that affected rope kits were incorrectly assembled, and could lead to the rope separating from the cone-shaped knob and causing the patient to fall back from the lift. The letters listed the specific quantities, order numbers and dates for the kits shipped to the customer, and requested the removal from use and return of the affected rope kits. Replacement rope kits were sent with the recall letters. Any questions were directed to Arjo''s Quality Department at 1-800-323-1245, ext. 6140.
Quantity in commerce
15 kits
Distribution
Nationwide, including Florida, Illinois, Indiana, Iowa, Kansas, Massachusetts, New York, Tennessee, Virginia and Wyoming.
Product codes and lots
Model KS1006, all kits shipped between July 2006 through September 2006