Z-0449-2007
Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX
JAS Diagnostics/Drew Scientific / initiated 2006-08-24 / Terminated
Drew Scientific Group PLC began a recall of Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX on . The reason FDA records is "Flags and alerts on the user interface will not automatically transfering to the LIS". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0449-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- JAS Diagnostics/Drew Scientific
- Product
- Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- Flags and alerts on the user interface will not automatically transfering to the LIS.
- FDA root cause
- Other
- Action
- Firm sent 'Product Notification #22' to all customers on 08/24/06. Users advised of potential risk to patient health and provided with instructions on how to eliminate risk by disabling the auto transfer feature and manually transfering data until a software patch is made available
- Quantity in commerce
- 299 units.
- Distribution
- Worldwide distribution --- including USA states of FL, MA, MN, ND, NM, NY, Puerto Rico and TX and countries of Austria, Bangladesh, Chile, China, Czech Republic, Egypt, France, Hong Kong, Italy, India, Iran, Korea, Mexico, Morocco, Phillipines, Portugal, Russia, Serbia and Montenegro, Spain, Syria, Thailand, Turkey, United Arab Emerates (Dubai), United Kingdom and Vietnam, West Indies and Yemen.-
- Product codes and lots
- All lots.