Z-0451-2007
Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX
JAS Diagnostics/Drew Scientific / initiated 2006-09-20 / Terminated
Drew Scientific Group PLC began a recall of Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX on . The reason FDA records is "System has given inaccurate readings when running samples that trigger an extra rinse due to very high WBC, RBC or Platlet counts". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0451-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- JAS Diagnostics/Drew Scientific
- Product
- Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- System has given inaccurate readings when running samples that trigger an extra rinse due to very high WBC, RBC or Platlet counts.
- FDA root cause
- Other
- Action
- Firm sent 'Technical Bulletin #39' to all customers on 09/20/06. Users advised of potential for inaccurate readings and provided with instructions on how to eliminate problem.
- Quantity in commerce
- 299 units.
- Distribution
- Worldwide distribution ---- including USA states of FL, MA, MN, ND, NM, NY, Puerto Rico and TX and countries of Austria, Bangladesh, Chile, China, Czech Republic, Egypt, France, Hong Kong, Italy, India, Iran, Korea, Mexico, Morocco, Phillipines, Portugal, Russia, Serbia and Montenegro, Spain, Syria, Thailand, Turkey, United Arab Emerates (Dubai), United Kingdom and Vietnam, West Indies and Yemen.
- Product codes and lots
- All lots.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
