Z-0451-2007

Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX

JAS Diagnostics/Drew Scientific / initiated 2006-09-20 / Terminated

Drew Scientific Group PLC began a recall of Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX on . The reason FDA records is "System has given inaccurate readings when running samples that trigger an extra rinse due to very high WBC, RBC or Platlet counts". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0451-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
JAS Diagnostics/Drew Scientific
Product
Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX.
Product code
GKZ Counter, Differential Cell
Reason for recall
System has given inaccurate readings when running samples that trigger an extra rinse due to very high WBC, RBC or Platlet counts.
FDA root cause
Other
Action
Firm sent 'Technical Bulletin #39' to all customers on 09/20/06. Users advised of potential for inaccurate readings and provided with instructions on how to eliminate problem.
Quantity in commerce
299 units.
Distribution
Worldwide distribution ---- including USA states of FL, MA, MN, ND, NM, NY, Puerto Rico and TX and countries of Austria, Bangladesh, Chile, China, Czech Republic, Egypt, France, Hong Kong, Italy, India, Iran, Korea, Mexico, Morocco, Phillipines, Portugal, Russia, Serbia and Montenegro, Spain, Syria, Thailand, Turkey, United Arab Emerates (Dubai), United Kingdom and Vietnam, West Indies and Yemen.
Product codes and lots
All lots.