Z-0224-2007

Roche Cobas driver (rchcob6i), analytical laboratory data interface

Data Innovations, LLC / initiated 2006-11-03 / Terminated / root cause: software

Data Innovations, Inc. began a recall of Roche Cobas driver (rchcob6i), analytical laboratory data interface on . The reason FDA records is "Software of modular driver may incorrectly report patient results as Quality Control Results". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0224-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Data Innovations, LLC
Product
Roche Cobas driver (rchcob6i), analytical laboratory data interface.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Software of modular driver may incorrectly report patient results as Quality Control Results
Root cause tag
software
FDA root cause
Software change control
Action
Data Innovations notified users by email listserv and by letter dated 10/25/06.
Quantity in commerce
56
Distribution
Nationwide and Canada.
Product codes and lots
Software version: v8.00.0001 and v7.00.0001