Z-0316-2007

Posterior Acetabular Retractor-Left with Light Cord Bracket, Model #BIO-1-1050-A, and Posterior Acetabular...

Whiteside Biomechanics, Inc. / initiated 2006-09-15 / Terminated

Whiteside Biomechanics began a recall of Posterior Acetabular Retractor-Left with Light Cord Bracket, Model #BIO-1-1050-A, and Posterior Acetabular Retractor-Right with Light Cord Bracket, Model... on . The reason FDA records is "The soft tissue pin in the retractor can break off and become embedded in the tissue". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0316-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Whiteside Biomechanics, Inc.
Product
Posterior Acetabular Retractor-Left with Light Cord Bracket, Model #BIO-1-1050-A, and Posterior Acetabular Retractor-Right with Light Cord Bracket, Model #BIO-1-1051-A, packaged 1 retractor (left or right retractor) per plastic bag. The firm name on the label is Whiteside Biomechanics, Inc., St. Louis, MO.
Product code
GAD Retractor
Reason for recall
The soft tissue pin in the retractor can break off and become embedded in the tissue.
FDA root cause
Other
Action
The recalling firm was notified of the problem via a telephone call from the surgeon on 9/15/06 and the remaining retractors were placed in quarantine by the hospital. The recalling firm requested return of the retractors. The recalling firm immediately visited the one remaining hospital on 9/15/06 and picked up their inventory of six retractors.
Quantity in commerce
10 retractors (5 right, 5 left)
Distribution
Distribution was made to two hospitals, one in MD and the other in MO.
Product codes and lots
Lot #P0835-1