BCI 3404 Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor. Smiths Medical PM, Inc. This device is...
Smiths Medical PM, Inc. began a recall of BCI 3404 Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor. Smiths Medical PM, Inc. This device is designed to provide full featured monitoring capabilities... on . The reason FDA records is "An error occurred at the board manufacturing site which may affect the following parameters: - An electrical noise observed on the SpO2 pleth waveform generating and displaying erratic heart rates. % SpO2 reading...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0300-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- BCI 3404 Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor. Smiths Medical PM, Inc. This device is designed to provide full featured monitoring capabilities in a tabletop design. The system features an ECG cable interface, an SpO2 probe interface, display of patient and waveform data via an EL panel, power status LED, and the function keypad area consisting of six keys (on/off, waveform/trend, alarm silence, menu/enter, up arrow & down arrow). The monitor has a serial port that is used for data communications to a printer or computer and for analog outputs.
- Product code
- DQA Oximeter
- Reason for recall
- An error occurred at the board manufacturing site which may affect the following parameters: - An electrical noise observed on the SpO2 pleth waveform generating and displaying erratic heart rates. % SpO2 reading displays dashes. - The ECG waveform may appear noisy, ECG readings appear accurate. -Respiration waveform may appear noisy and the respiration readings will be displayed as dashes.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- An Safety Action Bulletin, dated 11/03/2006 was sent to affected customers describing the issue, what parameters are affected on the monitor and provides informatin on how to have the monitor repaired. Smiths Medical MDPM requests that an acknowledgement form be returned to them.
- Quantity in commerce
- 207
- Distribution
- Worldwide, including USA, Puerto Rico, Argentina, Bolivia, Brunei, Chile, China, Columbia, Czech Republic, Germany, Greece, Hong Kong, India, Israel, Kuwait, Mexico, Netherlands, Poland, Romania, Singapore, Spain, Taiwan, and United Kingdom.
- Product codes and lots
- Catalog Number 3404001, serial number ranges: AI05070075 - AI05070099, AI05090000 - AI05090024, AI05110040 - AI05110059, AI05120000 - AI05120010, AI05120013 - AI05120024, AI06010035 - AI06010049, AI06040000 - AI06040024, AI06060000 - AI06060014, AI06060016 - AI06060024, AI06070035 - AI06070041, AI06070043 - AI06070045. Catalog Number 3401001S, serial number range: AI05110000 - AI05110039