DRLock Volar Plate Screw System, Model #DVR-900. The product is a kit contained in an aluminum tray case that...
OrthoHelix Surgical Designs Inc began a recall of DRLock Volar Plate Screw System, Model #DVR-900. The product is a kit contained in an aluminum tray case that contains the following: IMPLANTS: 6 stainless... on . The reason FDA records is "The 48 distal locking-screws (ranging from 10-24mm) contained in the kit are not within established specifications, specifically the minor diameter of the screws are undersized". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0349-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- TORNIER Inc.
- Product
- DRLock Volar Plate Screw System, Model #DVR-900. The product is a kit contained in an aluminum tray case that contains the following: IMPLANTS: 6 stainless steel plates (2-lefts, 2-rights, and 1-right long and 1-right left), 48 distal locking-screws (ranging from 10-24mm), 32 distal locking pegs, 36 proximal locking screws, 24 proximal non-locking screws. INSTRUMENTS: 2 Bending Pliers*, 1 Bone Reduction Forcep*, 2 Hohmann Retractors *, 1 Ratchet Handle*, 1 Periosteal Elevator*, 1 Dingman Elevator *, 1 K & K Elevator *, four Driver tips *, five drill guides*, 2-1.6 Drill Bits *, 2-2.3 Drill Bits *, 1 Plate Holder Verbruge *, 2 Depth Gages *, 6 K wires.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The 48 distal locking-screws (ranging from 10-24mm) contained in the kit are not within established specifications, specifically the minor diameter of the screws are undersized.
- FDA root cause
- Other
- Action
- The recalling firm contacted all consignees by phone on or about 8/3/06.
- Quantity in commerce
- 180
- Distribution
- The product was distributed to 5 consignees located in the U.S.: Archway Medical, Columbia, SC; Gulf Coast Surgical, Stafford, TX; Vaugham Medical, Wheat Ridge, CO; Belemy Medical, Cleveland, OH; and Crystal Clinic, Akron, OH.
- Product codes and lots
- Tray kit (case) set number: ''BAA'', ''BAB'', ''BAC'', ''BAD'', ''BAE'', ''BAF'', ''BAH'', and ''BAJ''. (The distal locking-screws contained in the kit are identified with lot numbers S104 and S200)