Z-0260-2007

Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by...

Abbott Medical / initiated 2006-11-17 / Terminated / root cause: labeling and UDI

Smiths Medical ASD, Inc. began a recall of Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S on . The reason FDA records is "Mislabeled as to size: The product is 47.5 mm in length, but the label lists the length as 4.75 mm". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0260-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Custom 5.5 mm Bivona Uncuffed Hyperflex Tracheostomy Tube, I.D. 5.5 mm, O.D. 8.0 mm, length 4.75 mm, Mfrd by Smiths Medical Critical Care; REF HU05JS55NNA112S.
Product code
JOH Tube Tracheostomy And Tube Cuff
Reason for recall
Mislabeled as to size: The product is 47.5 mm in length, but the label lists the length as 4.75 mm.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
The consignee was notified of the problem via recall letter sent on 11/17/06 and given the options of continuing to use the product or of returning it.
Quantity in commerce
5
Distribution
New York-single hospital consignee
Product codes and lots
Lot CL47189.