Z-0504-2007
Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden
Stille Surgical Inc. / initiated 2006-11-09 / Terminated
Stille-Sonesta, Inc began a recall of Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden on . The reason FDA records is "Bolts may back out of their joints causing the back section of the table to tilt". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0504-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stille Surgical Inc.
- Product
- Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.
- Product code
- IXR Table, Radiographic, Tilting
- Reason for recall
- Bolts may back out of their joints causing the back section of the table to tilt.
- FDA root cause
- Other
- Action
- Notifications letters sent to consignees on 11/20/06. Users are instructed to check the bolts located under the table. The Notice includes instructions with pictures for checking the bolts and tightening them if necessary. Stille AB has developed a kit for replacing the bolt that is involved and will send a service engineer out to the site to do the replacement.
- Quantity in commerce
- 66 units.
- Distribution
- Nationwide.
- Product codes and lots
- Serial Numbers: 12-70000 - 12-70017, 00010, 00012, 00013 - 00022, 00024 - 00027, 00030 - 00034, 00036 - 00040, 00043 - 00044, 00045 - 00049, 00053 - 00054, 00057 - 00065, 00067 - 00069, 00071. Outside U.S. Serial Numbers: 12-70018 - 12-70019, 23-70000 - 23-70009, 23-70008, 23-700010, 00011, 00023, 00028 - 00029, 00050 - 00052, 00055 - 00056, 00066, 00072, 00073, 00074.