Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden

Stille Surgical Inc. / initiated 2006-11-09 / Terminated

Stille-Sonesta, Inc began a recall of Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden on . The reason FDA records is "Bolts may back out of their joints causing the back section of the table to tilt". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0504-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stille Surgical Inc.
Product
Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.
Product code
IXR Table, Radiographic, Tilting
Reason for recall
Bolts may back out of their joints causing the back section of the table to tilt.
FDA root cause
Other
Action
Notifications letters sent to consignees on 11/20/06. Users are instructed to check the bolts located under the table. The Notice includes instructions with pictures for checking the bolts and tightening them if necessary. Stille AB has developed a kit for replacing the bolt that is involved and will send a service engineer out to the site to do the replacement.
Quantity in commerce
66 units.
Distribution
Nationwide.
Product codes and lots
Serial Numbers:  12-70000 - 12-70017, 00010, 00012, 00013 - 00022, 00024 - 00027, 00030 - 00034, 00036 - 00040, 00043 - 00044, 00045 - 00049, 00053 - 00054, 00057 - 00065, 00067 - 00069, 00071.  Outside U.S. Serial Numbers: 12-70018 - 12-70019, 23-70000 - 23-70009, 23-70008, 23-700010, 00011, 00023, 00028 - 00029, 00050 - 00052, 00055 - 00056, 00066, 00072, 00073, 00074.