Z-0301-2007
Infinity Docking Station (IDS) Delta and Kappa Series patient Monitors. Model numbers 4715319, 5206110...
Draeger, Inc. / initiated 2006-11-17 / Terminated
Draeger Medical, Inc. began a recall of Infinity Docking Station (IDS) Delta and Kappa Series patient Monitors. Model numbers 4715319, 5206110, 5732388, and 7489375 on . The reason FDA records is "Top cover of assembly may separate. The situation can result in the monitor falling, which could result in an injury to the clinician or patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0301-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- Infinity Docking Station (IDS) Delta and Kappa Series patient Monitors. Model numbers 4715319, 5206110, 5732388, and 7489375.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Top cover of assembly may separate. The situation can result in the monitor falling, which could result in an injury to the clinician or patient.
- FDA root cause
- Other
- Action
- The recaling firm issued a recall letter to the US customers 11/17/2006 via certified mail. An authorized service representative will inspect each potentially affected unit. If a unit is found to have an affected docking station, the unit will be repaired or replaced free or charge. Notification to international customers will occur through the international subsidiaries.
- Quantity in commerce
- 23,345 units - 4336 US
- Distribution
- Nationwide
- Product codes and lots
- Model numbers 4715319, 5206110, 5732388, and 7489375.