Davol Salute Fixation System, 38cm Reusable Shaft (Laparoscopic) with Handle. To attach mesh to the abdominal...
Davol, Inc., Sub. C. R. Bard, Inc. began a recall of Davol Salute Fixation System, 38cm Reusable Shaft (Laparoscopic) with Handle. To attach mesh to the abdominal wall. Reference Number 0113036 or... on . The reason FDA records is "Inadequate directions for use in the Salute System. Labeling -Revised to provide updated instructions for use and proper servicing". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0374-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Davol Inc., Sub. C.R. Bard, Inc.
- Product
- Davol Salute Fixation System, 38cm Reusable Shaft (Laparoscopic) with Handle. To attach mesh to the abdominal wall. Reference Number 0113036 or 9113036(refurbished)
- Product code
- GDW Staple, Implantable
- Reason for recall
- Inadequate directions for use in the Salute System. Labeling -Revised to provide updated instructions for use and proper servicing
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Davol notified users by a first letter on 12/27/06 providng a introductory letter explaining the Enhanced Servicing Program and changes made to the IFU (re: servicing, tip inspection, end of wire, and test fire protocol). A second letter will issue to accounts to identifying the serial number to be serviced and returned to Davol.
- Quantity in commerce
- 4787 (includes both types)
- Distribution
- Worldwide, including USA, Austria, Italy, Israel, Finland, Switzerland, Portugal, Japan, France, Spain, Greece, Belgium, Germany, Sweden, Holland, Denmark, UK, and South Africa.
- Product codes and lots
- All serial numbers