Disposable components of the superDimension Bronchus System: Locatable Guide Kit (AKI00017-01) and Extended...
SuperDimension, Inc. began a recall of Disposable components of the superDimension Bronchus System: Locatable Guide Kit (AKI00017-01) and Extended Working Channel (AKI00041-01) on . The reason FDA records is "The action is being taken in response to certain incidents that have been reported to the company, one of which concerned an adverse event in which a patient incurred a pneumothorax during a procedure in which a...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0667-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Disposable components of the superDimension Bronchus System: Locatable Guide Kit (AKI00017-01) and Extended Working Channel (AKI00041-01).
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The action is being taken in response to certain incidents that have been reported to the company, one of which concerned an adverse event in which a patient incurred a pneumothorax during a procedure in which a superDimension Bronchus System was used. The biopsy results for the patient were taken from a different area than the confirmed navigation site that was indicated by the system.
- FDA root cause
- Other
- Action
- An Urgent Medical Device Field Action letter, dated 11/30/2006, was sent to medical facilities who have the superDimension Bronchus System installed. The letter infoms the facility to immediately stop using the Bronchus System, and move all inventory of the disposable system components to a controlled quarantine area. A superDimension representative will contact the consignees to schedule a time to retrieve the quarantined inventory, at which time a verification form will be completed.
- Quantity in commerce
- Up to 2,000 each Locatable Guide Kit and Extended Working Channel
- Distribution
- Worldwide, including USA, Austria, Belgium, Canada, France, Germany, Hungary, Israel, Italy, Netherlands, South Africa, Spain, Switzerland, Turkey, and United Kingdom.
- Product codes and lots
- All codes of disposable system components (Locatable Guide and Extended Working Channel).