Fisher & Paykel Oracle Oral Mask, Part Numbers: 900HC451, 900HC451T, HC451A, HC452A & HC452F. CPAP masks
Fisher & Paykel Healthcare Inc began a recall of Fisher & Paykel Oracle Oral Mask, Part Numbers: 900HC451, 900HC451T, HC451A, HC452A & HC452F. CPAP masks on . The reason FDA records is "Plastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0973-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fisher & Paykel Healthcare, Inc.
- Product
- Fisher & Paykel Oracle Oral Mask, Part Numbers: 900HC451, 900HC451T, HC451A, HC452A & HC452F. CPAP masks.
- Product code
- BZD Ventilator, Non-Continuous (Respirator)
- Reason for recall
- Plastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and there is an extremely remote possibility that the broken tab may then enter the patient's nasal cavity or mouth. Masks manufactured since April 2006 do not feat
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Fisher & Paykel Healthcare Inc. commenced sending out letters by mail to consignees the week commencing 4 December 2006 advising consignees to quarantine and destroy affected product. A faxback form was also be provided for consignees to advise Fisher & Paykel Healthcare of quarantined and destroyed stock. The firm will supply consignees with replacement product as determined from completed faxback information. The firm will collate responses from consignees and will follow up with a letter to consignees who do not respond within a month of initial mail-out. The firm will then make telephone contact with any consignees who do not respond within a month of the second mail-out. Fisher & Paykel Healthcare Inc. estimated the completion of this voluntary recall by April 2007.
- Quantity in commerce
- 815,239 total for all models, all lots
- Distribution
- Nationwide
- Product codes and lots
- Lot numbers for Part Numbers: 900HC451 (lot #s: 011001 to 011016), 900HC451T (lot #s: 011024 to 020517), HC451A (lot #s: 020523 to 040907), HC452A (lot #s: 031215 to 060503) & HC452F (lot #s: 040719 to 041102)
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
