Z-0423-2007
OsSatura TCP (pure tricalcium phosphate) 3mm granules packaged in glass jars, Part Numbers: 05-6010-050 5cc...
Teknimed SAS / initiated 2006-08-31 / Terminated / root cause: labeling and UDI
Teknimed SA began a recall of OsSatura TCP (pure tricalcium phosphate) 3mm granules packaged in glass jars, Part Numbers: 05-6010-050 5cc, 05-6010-100 10cc, 05-6010-150 15cc & 05-6010-300... on . The reason FDA records is "The firm has determined an incorrect Instructions for Use (IFU) was packaged within the product box". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0423-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teknimed SAS
- Product
- OsSatura TCP (pure tricalcium phosphate) 3mm granules packaged in glass jars, Part Numbers: 05-6010-050 5cc, 05-6010-100 10cc, 05-6010-150 15cc & 05-6010-300 30cc
- Product code
- MQV Filler, Bone Void, Calcium Compound
- Reason for recall
- The firm has determined an incorrect Instructions for Use (IFU) was packaged within the product box.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Teknimed notified IsoTis OrthoBiologics via electronic recall letter dated August 31, 2006. IsoTis was instructed to notify their consignees of this field correction and they were asked to be responsible for identifying and making any corrections necessary at the medical user level. Teknimed supplied the correct IFU to IsoTis so that they could perform a field correction of the affected product. October 1, 2006 IsoTis OrthoBiologics sent their consignees a recall notification that included the IsoTis notification letter, a copy of the Teknimed recall letter, IsoTis Field Correction Instruction Sheet and replacement TCP IFU's for the affected lots.
- Quantity in commerce
- 1681 units
- Distribution
- Nationwide
- Product codes and lots
- Batch numbers shipped to US: 82/0415, 82/04474, 82/05062, 82A/05454, 82/04172, 82/04264, 82/04274, 82/04472, 82/04511, 82/05052, 82A/0526, 82A/05404, 82A/0619, 82/04182, 82/04204, 82/04252, 82A/05452, 82A/06014, 82A/0620, 82/04194, 82/04242, 82/04492 & 82A/06024