ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; Temperature Controller...
Abbott Laboratories, Inc began a recall of ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; Temperature Controller Board Part #78560-109; distributed by Abbott... on . The reason FDA records is "The re-designed ARCHITECT Temperature Controller Board has been identified as being susceptible to electromagnetic interference (EMI) in the laboratory. This can cause a board reset condition, stopping temperature...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0415-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; Temperature Controller Board Part #78560-109; distributed by Abbott Laboratories, Irving, TX.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- The re-designed ARCHITECT Temperature Controller Board has been identified as being susceptible to electromagnetic interference (EMI) in the laboratory. This can cause a board reset condition, stopping temperature control function of the board. Situation does not stop analyzer operation and does not generate an error condition alerting operator when condition occurs.
- FDA root cause
- Other
- Action
- Firm contacted consignees by telephone and fax with instructions to verify proper functioning of the Temperature Controller Board starting December 13, 2006.
- Quantity in commerce
- 57 units.
- Distribution
- Worldwide, including USA, Argentina, Australia, Brazil, Canada, Chile, Columbia, Dominican Republic, Germany, Hong Kong, Jamaica, Japan, Korea, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Taiwan, Thailand, Trinidad, Tobago, Uruguay, and Venezuela.
- Product codes and lots
- Serial numbers: 1202790, 1202804, 1202805, 1202806, 1202807, 1202808, 1202809, 1202810, 1202811, 1202812, 1202813, 1202814, 1202815, 1202816, 1202817, 1202818, 1202819, 1202820, 1202821, 1202822, 1202823, 1202824, 1202825, 1202826, 1202827, 1202828, 1202829, 1202830, 1202831, 1202832, 1202833, 1202834, 1202835, 1202836, 1202837, 1202838, 1202839, 1202840, 1202841, 1202842, 1202843, 1202844, 1202845, 1202846, 1202847, 1202848, 1202849, 1202850, 1202851, 1202852, 1202853, 1202854, 1202855, 1202856, 1202857, 1202858, 1202859.