Z-0361-2007
Salicylate-SL, in vitro diagnostic, Cat. No.511-20, Lot No. 29892
Sekisui Diagnostics P.E.I. Inc. / initiated 2006-12-05 / Terminated
Diagnostic Chemicals, Ltd. began a recall of Salicylate-SL, in vitro diagnostic, Cat. No.511-20, Lot No. 29892 on . The reason FDA records is "Upon subsequent testing on different instruments (as required by other customers), DCL detected a non-linear response on certain analyzers. A decision was made to recall the product due to the fact that DCL customers...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0361-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sekisui Diagnostics P.E.I. Inc.
- Product
- Salicylate-SL, in vitro diagnostic, Cat. No.511-20, Lot No. 29892
- Product code
- DKJ Colorimetry, Salicylate
- Reason for recall
- Upon subsequent testing on different instruments (as required by other customers), DCL detected a non-linear response on certain analyzers. A decision was made to recall the product due to the fact that DCL customers sell the product to different end users, DCL cannot be certain which instruments the end users employ in their laboratories
- FDA root cause
- Other
- Action
- For product shipped by DCL to one customer notification letter dated 12/5/06 sent out to customer to return/discard product. Replacement will be sent to customer. For product shipped to DCL (USA) warehouse notification sent to discard kits remaining in stock. For product remaining at DCL Canada in-house notification sent out to discard in-house kits.
- Quantity in commerce
- 55 kits shipped to one customer
- Distribution
- PA
- Product codes and lots
- Lot 29892