Philips Integris Allura x-ray systems angiographic - Allura 9 F/C (floor and ceiling models). A mono-plane...
Philips Medical Systems North America Co. Phillips began a recall of Philips Integris Allura x-ray systems angiographic - Allura 9 F/C (floor and ceiling models). A mono-plane analog Allura with a 9' Image Intensifier on . The reason FDA records is "Movement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0366-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Integris Allura x-ray systems angiographic - Allura 9 F/C (floor and ceiling models). A mono-plane analog Allura with a 9' Image Intensifier
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- Movement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.
- FDA root cause
- Other
- Action
- On 12/14/05 and 12/20/06 the firm issued a letter dated 12/12/06 to their customers. The letter explains the problem and states that the firm will be performing mandatory upgrades to correct the situation. Consignees will be contacted to arrange for service.
- Quantity in commerce
- 140 devices total
- Distribution
- Units were distributed to hospitals and medical centers nationwide.
- Product codes and lots
- site numbers 534928, 545964, 532207, 545365, 538387, 543397, 543398, 543399, 546572, 544581, 544728, 548147, 547005, 539456, 543953, 547234, 547364, 545838, 545839, 545840, 544395, 547952, 530393, 545799, 540790, 547431, 544122, 541659, 546060, 543433, 545604, 543198, 545921, 545640, 545966, 541590, 540456, 546201, 541572, 544130, 544635, 534665, 542963, 535303, 541783, 544631, 540930, 544610, 544506, 546456, 544989, 544990, 544992, 543984, 543759, 546906, 543873, 543087, 543108, 547279, 546908, 544750, 545518, 545052, 545539, 543799, 543411, 547660, 543229, 544371, 548288, 546668, 545714, 537604, 541046, 544513, 545583, 540808, 545280, 548039, 544489, 544120, 540275, 540284, 540952, 544637, 547855, 545941, 545753, 541597, 545025, 536030, 544543, 546693, 548140, 546025, 545752, 547976, 543971, 543972, 539634, 545749, 546119, 545750, 547642, 544433, 545069, 545073, 547777, 547321, 547628, 545733, 547392, 545293, 545296, 547908, 541880, 543386, 546013, 548061, 545282, 533398, 541588, 541589, 540374, 545357, 542486, 545443, 546012, 545090, 545037, 538432, 532226, 540407, 103281, 104950, 531926, 104953, 530653, 520630