Jostra Heater Cooler Unit HCU30, system version 1.02-Low voltage machine

Maquet Cardiovascular Us Sales, LLC / initiated 2006-11-10 / Terminated

MAQUET Inc. began a recall of Jostra Heater Cooler Unit HCU30, system version 1.02-Low voltage machine on . The reason FDA records is "The HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0350-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Jostra Heater Cooler Unit HCU30, system version 1.02-Low voltage machine
Product code
DWC Controller, Temperature, Cardiopulmonary Bypass
Reason for recall
The HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or cardioplegia sufficiently.
FDA root cause
Other
Action
Letters were mailed to customers on 11/10/2006 by Federal Express.
Quantity in commerce
13 units
Distribution
Hospitals in CA, NC, FL, WY, NY and WV.S
Product codes and lots
Article number: 0704629 Serial numbers: 101006, 101007, 101008, 101009, 101010, 101011, 101013, 101015, 101016, 101018, 101019, 101020, 101022, 101023, 101024.