Z-0350-2007
Jostra Heater Cooler Unit HCU30, system version 1.02-Low voltage machine
Maquet Cardiovascular Us Sales, LLC / initiated 2006-11-10 / Terminated
MAQUET Inc. began a recall of Jostra Heater Cooler Unit HCU30, system version 1.02-Low voltage machine on . The reason FDA records is "The HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0350-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Jostra Heater Cooler Unit HCU30, system version 1.02-Low voltage machine
- Product code
- DWC Controller, Temperature, Cardiopulmonary Bypass
- Reason for recall
- The HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or cardioplegia sufficiently.
- FDA root cause
- Other
- Action
- Letters were mailed to customers on 11/10/2006 by Federal Express.
- Quantity in commerce
- 13 units
- Distribution
- Hospitals in CA, NC, FL, WY, NY and WV.S
- Product codes and lots
- Article number: 0704629 Serial numbers: 101006, 101007, 101008, 101009, 101010, 101011, 101013, 101015, 101016, 101018, 101019, 101020, 101022, 101023, 101024.