Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0603-2007

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required)...

Terumo Cardiovascular Systems Corporation / initiated 2005-08-01 / Terminated

Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the... on . The reason FDA records is "The power supply may fail to function due to a defective capacitor. (Note; this device incorporates a redundant separate power supply not affected by this capacitor)". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0603-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
The power supply may fail to function due to a defective capacitor. (Note; this device incorporates a redundant separate power supply not affected by this capacitor)
FDA root cause
Other
Action
The power supplies were replaced as complaints were received between August 2005 and February 2006. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their units will be corrected when replacement components become available.
Quantity in commerce
534 of both model numbers
Distribution
Worldwide Distribution: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Jordan, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore and South Africa.
Product codes and lots
Serial numbers 0006 through 0066 and 0100 through 0187 may be affected. Serial numbers 0046 and 0160 have been corrected.