Z-0417-2007
Access Immunoassay Systems Thyroglobulin Antibody Reagent Kit, Part Number 33890
Siemens Healthcare Diagnostics Inc. / initiated 2006-10-26 / Terminated / root cause: manufacturing process
Beckman Coulter Inc began a recall of Access Immunoassay Systems Thyroglobulin Antibody Reagent Kit, Part Number 33890 on . The reason FDA records is "Low end imprecision affecting both S0 calibrators and patient samples which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0417-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Access Immunoassay Systems Thyroglobulin Antibody Reagent Kit, Part Number 33890
- Product code
- DDC Thyroglobulin, Antigen, Antiserum, Control
- Reason for recall
- Low end imprecision affecting both S0 calibrators and patient samples which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- A Product Corrective Action (PCA) letter was mailed on Oct 27, 2006 to all Thyroglobulin Antibody customers. They are informed that there has been an increase in customer feedback regarding low end imprecision affecting both S0 calibrators and patient samples, which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay. The letter informs the customer that the firm recommends that they use an alternate method for testing TgAb. (Letter was sent by US mail). Customers are asked to complete and return the enclosed response form within 10 days.
- Quantity in commerce
- 474 units
- Distribution
- Nationwide and Canada
- Product codes and lots
- 518729 (expires 10/31/06); 520617 (expires 12/31/06); 612764 (expires 12/31/06); 613481 (expires 3/31/07); 614541 (expires 5/31/07); 616787 (expires 8/31/07)