Z-0521-2007
Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker...
Stryker Craniomaxillofacial Division / initiated 2006-12-22 / Terminated
Stryker Leibinger USA began a recall of Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407 on . The reason FDA records is "Metal shavings come off of the plug as a result of friction between the plug and the cutting cylinder, and may be mixed with the milled bone chips". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0521-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Craniomaxillofacial Division
- Product
- Stryker Tom Tessier Osseous Microtome: Plug for Bone Mill Systems (Parts 01-15400 and 01-15401), Stryker Leiblinger Inc., Kalamazoo, MI; Part 01-15407.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Metal shavings come off of the plug as a result of friction between the plug and the cutting cylinder, and may be mixed with the milled bone chips.
- FDA root cause
- Other
- Action
- Consignees were notified via recall letter dated 12/22/06 which instructed them to quarantine the product and to contact the firm for further instructions.
- Quantity in commerce
- 26
- Distribution
- Nationwide.
- Product codes and lots
- Lots G4C00F72HN, G5H00F72HN, G7T00F72HN, G8W00F72HN and G9700F72HN.