Z-0372-2007

Ventilator System SERVO-i V3.1 PC1777 Maquet Critical Care

Maquet Cardiovascular Us Sales, LLC / initiated 2006-05-23 / Terminated

MAQUET Inc. began a recall of Ventilator System SERVO-i V3.1 PC1777 Maquet Critical Care on . The reason FDA records is "Reported failures of the SERVO-i ventilators user interface system causing technical errors and increased fault frequencies. Switches may be activated unintentionally, or a function not activate when the switch is...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0372-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Ventilator System SERVO-i V3.1 PC1777 Maquet Critical Care
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Reported failures of the SERVO-i ventilators user interface system causing technical errors and increased fault frequencies. Switches may be activated unintentionally, or a function not activate when the switch is selected. Direct access knobs may not function or change the parameter in the wrong direction compared to how they are turned.
FDA root cause
Other
Action
Maquet sent an ''Extended Warranty'' letter on May 23, 2006. On March 13, 2007, Maquet issued a Recall Notification letter to all customers stating that they will replace all PC1777 boards in the serial number range 2400-3300, which have not previously been replaced. A safety alert letter was included outlining the failure indication and effect for the user of the failures, as well as, how to identify whether the Servo-i unit has shown any failure indications.
Quantity in commerce
1, 970 units
Distribution
Worldwide
Product codes and lots
Article # 64-87-800 Serial numbers 02400 - 03300