Z-0372-2007
Ventilator System SERVO-i V3.1 PC1777 Maquet Critical Care
Maquet Cardiovascular Us Sales, LLC / initiated 2006-05-23 / Terminated
MAQUET Inc. began a recall of Ventilator System SERVO-i V3.1 PC1777 Maquet Critical Care on . The reason FDA records is "Reported failures of the SERVO-i ventilators user interface system causing technical errors and increased fault frequencies. Switches may be activated unintentionally, or a function not activate when the switch is...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0372-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Ventilator System SERVO-i V3.1 PC1777 Maquet Critical Care
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Reported failures of the SERVO-i ventilators user interface system causing technical errors and increased fault frequencies. Switches may be activated unintentionally, or a function not activate when the switch is selected. Direct access knobs may not function or change the parameter in the wrong direction compared to how they are turned.
- FDA root cause
- Other
- Action
- Maquet sent an ''Extended Warranty'' letter on May 23, 2006. On March 13, 2007, Maquet issued a Recall Notification letter to all customers stating that they will replace all PC1777 boards in the serial number range 2400-3300, which have not previously been replaced. A safety alert letter was included outlining the failure indication and effect for the user of the failures, as well as, how to identify whether the Servo-i unit has shown any failure indications.
- Quantity in commerce
- 1, 970 units
- Distribution
- Worldwide
- Product codes and lots
- Article # 64-87-800 Serial numbers 02400 - 03300