Z-0805-2007
GE LightSpeed Xtra / GE LightSpeed RT Pro 16, Model 2374681-6, CT Scanner systems
GE Healthcare / initiated 2006-12-21 / Terminated
GE Healthcare began a recall of GE LightSpeed Xtra / GE LightSpeed RT Pro 16, Model 2374681-6, CT Scanner systems on . The reason FDA records is "Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0805-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE LightSpeed Xtra / GE LightSpeed RT Pro 16, Model 2374681-6, CT Scanner systems.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.
- FDA root cause
- Other
- Action
- A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
- Quantity in commerce
- 21
- Distribution
- Worldwide Distribution: USA in the following states and locations: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, Guam and Puerto Rico. Foreign countries include ; Albania, Algeria, Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Canada, Chile, China, Columbia, Czechoslovakia, Denmark, Finland, France, Germany, Great Britain, Greece, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Reunion Island, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Venezuela, and Vietnam
- Product codes and lots
- All serial numbers