Z-0520-2007

Quantum TTC Biliary Balloon Dilation Catheter, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road...

Wilson-Cook Medical Inc. / initiated 2007-01-24 / Terminated / root cause: labeling and UDI

Cook Endoscopy began a recall of Quantum TTC Biliary Balloon Dilation Catheter, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105 on . The reason FDA records is "The graphic label is incorrect, it lists 6mm/ 18FR. The product label is correct, it lists 10mm/ 30FR". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0520-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Wilson-Cook Medical Inc.
Product
Quantum TTC Biliary Balloon Dilation Catheter, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105
Product code
KNQ Dilator, Esophageal
Reason for recall
The graphic label is incorrect, it lists 6mm/ 18FR. The product label is correct, it lists 10mm/ 30FR.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Consignees were notified by UPS certified mail and email on/about 01/31/2007. They were instructed to return any affected product to Cook Endoscopy. A response form was attached to be completed and returned to Cook Endoscopy.
Quantity in commerce
12 units
Distribution
Worldwide, including NY and TX in USA, Australia, Canada and India.
Product codes and lots
Order number QBD-10X3, Lot number W2287708