Z-0499-2007
Anaerotest;Microbiology , 50 test strips 1.15112 Store dry and tightly closed. Store at + 15 C to +25C Merck...
EMD Millipore Corporation / initiated 2006-08-17 / Terminated
EMD Chemicals Inc began a recall of Anaerotest;Microbiology , 50 test strips 1.15112 Store dry and tightly closed. Store at + 15 C to +25C Merck IVD 64271 Darmstadt, Germany on . The reason FDA records is "Lot # OC553145 of Anaerotest failed to remain in specification. QC indicator shows an anaerobic environment even if oxygen is present". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0499-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- EMD Millipore Corporation
- Product
- Anaerotest;Microbiology , 50 test strips 1.15112 Store dry and tightly closed. Store at + 15 C to +25C Merck IVD 64271 Darmstadt, Germany
- Product code
- J
- Reason for recall
- Lot # OC553145 of Anaerotest failed to remain in specification. QC indicator shows an anaerobic environment even if oxygen is present
- FDA root cause
- Other
- Action
- A Recall Letter was issued by EMD Chemicals on 8/17/2006 to discontinue product usage and for product return.
- Quantity in commerce
- 800 packages of 50 strips each.
- Distribution
- Product was only sent to one US customer: Becton Dickinson, Cayey. Pureto Rico.
- Product codes and lots
- Item # 1.15112.0001 Lot # OC553145