Z-0923-2007
Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040
Terumo Cardiovascular Systems Corporation / initiated 2004-04-01 / Terminated
Terumo Cardiovascular Systems Corp began a recall of Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040 on . The reason FDA records is "A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0923-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease.
- FDA root cause
- Other
- Action
- The firm is visiting consignees and making corrections as complaints are received. These corrections, to date, have been made between 04/04 and 09/06. The firm will correct other units as complaints are received and issued a Device Correction Letter dated 6/6 /07 notifying all users of the problem and appropriate responses should it occur.
- Quantity in commerce
- 525 total for both models
- Distribution
- Worldwide: USA, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Singapore and United Arab Emirates.
- Product codes and lots
- Serial numbers 0031 through 0324 are potentially affected by the problem. Serial numbers 0092, 0128, 0152, 0223, 0226, 0237, 0246, 0260 and 0296 have been corrected.