Z-0582-2007
BD Integra 1ml Insulin Syringe with Retracting Precision Guide Needle 1 ml 29g 1/2'' (0.33mm x 13 mm)...
Becton Dickinson & Co / initiated 2007-01-27 / Terminated
Becton Dickinson & Company began a recall of BD Integra 1ml Insulin Syringe with Retracting Precision Guide Needle 1 ml 29g 1/2'' (0.33mm x 13 mm) Re-Order No. 305282 U-100 insulin Lot 6073436 on . The reason FDA records is "The needle becomes detached from the hub/syringe". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0582-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Integra 1ml Insulin Syringe with Retracting Precision Guide Needle 1 ml 29g 1/2'' (0.33mm x 13 mm) Re-Order No. 305282 U-100 insulin Lot 6073436
- Product code
- FMF Syringe, Piston
- Reason for recall
- The needle becomes detached from the hub/syringe
- FDA root cause
- Other
- Action
- Recall notification letters were sent via UPS second day air on January 27, 2007. Distributors were asked to provide end user contact information to recalling firm.
- Quantity in commerce
- 247,824 units
- Distribution
- Nationwide
- Product codes and lots
- Lot # 6073436 Ref (Reorder number) 305282)