Z-0575-2007
StarClose Vascular Closure System (Clip Applier & Exchange System), Catalog Number: 14677; Product is...
Abbott Vascular / initiated 2007-01-31 / Terminated
Abbott Vascular, Inc. began a recall of StarClose Vascular Closure System (Clip Applier & Exchange System), Catalog Number: 14677; Product is manufactured by: Abbott Vascular Inc., 400 Saginaw Drive... on . The reason FDA records is "Premature release of the vessel locator wings, which stabilize the device prior to clip deployment, will result in no hemostasis". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0575-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Vascular
- Product
- StarClose Vascular Closure System (Clip Applier & Exchange System), Catalog Number: 14677; Product is manufactured by: Abbott Vascular Inc., 400 Saginaw Drive, Redwood City, CA 94063
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Premature release of the vessel locator wings, which stabilize the device prior to clip deployment, will result in no hemostasis.
- FDA root cause
- Other
- Action
- On 1/31/07, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned to its sales rep.
- Quantity in commerce
- 97,952 units
- Distribution
- Worldwide distribution, including USA, Puerto Rico, and 410 other countries. There were no shipments to Canada and Mexico.
- Product codes and lots
- Lot Numbers starting from 43034-6H to 47139-6H