Z-0766-2007

Bausch & Lomb Millennium TSV 25 Ga. High Speed Posterior Vitrectomy Pack, Catalog No: CX4925, containing...

Bausch & Lomb Incorporated / initiated 2007-02-09 / Terminated

Bausch & Lomb Inc began a recall of Bausch & Lomb Millennium TSV 25 Ga. High Speed Posterior Vitrectomy Pack, Catalog No: CX4925, containing 1/25-ga. cutter along with accessories, RX. The firm... on . The reason FDA records is "The cutter probe tip breaks during use". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0766-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bausch & Lomb Incorporated
Product
Bausch & Lomb Millennium TSV 25 Ga. High Speed Posterior Vitrectomy Pack, Catalog No: CX4925, containing 1/25-ga. cutter along with accessories, RX. The firm names on the label are Bausch & Lomb Incorporated, Rochester, NY, and Bausch & Lomb Incorporated, Kingston-upon-Thames, UK.
Product code
HQE Instrument, Vitreous Aspiration And Cutting, Ac-Powered
Reason for recall
The cutter probe tip breaks during use.
FDA root cause
Other
Action
On 2/12/07, express mail letters were addressed to operating room supervisors to notify them of the recall. Additional E-mail notifications were issued on 2/14/07 to doctors and letters were addressed to them on 2/20/07 to make them aware of the recall, although they do not directly receive the product. The communications instructed them to not use the affected product and to return it, along with the business reply card.
Quantity in commerce
13,685 units of Catalog No: CX4925
Distribution
Class II Recall - Worldwide Distribution --- USA including states of MO, NC, FL, CA, and TX and countries of Canada, China, Japan, Australia, Hong Kong, India, Singapore, Korea, Taiwan, Thailand, Argentina, Brazil, Mexico, and The Netherlands
Product codes and lots
All lots