HTA Console System, 230 Volt , Italian, Product Number: 56001 IO (HTA system is a software-controlled...
Boston Scientific Corporation began a recall of HTA Console System, 230 Volt , Italian, Product Number: 56001 IO (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system) on . The reason FDA records is "Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0691-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- HTA Console System, 230 Volt , Italian, Product Number: 56001 IO (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)
- Product code
- MNB Device, Thermal Ablation, Endometrial
- Reason for recall
- Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment
- FDA root cause
- Other
- Action
- Boston Scientific notiifed accounts by letter dated January 24, 2007 via federal Express Mail. Accounts were requested to examine inventory, remove and return product.
- Quantity in commerce
- 2 units
- Distribution
- Worldwide-USA, UK, Denmark, Italy, and France
- Product codes and lots
- Unit Numbers: IN0339JC, IN0340JC,