Z-0638-2007
VIASYS VELA Ventilator, Catalog Number-16186; Viasys Respiratory Care, Inc
ZOLL Palm Springs / initiated 2007-01-31 / Terminated
Viasys Respiratory Care, Inc.dba Bird Products began a recall of VIASYS VELA Ventilator, Catalog Number-16186; Viasys Respiratory Care, Inc on . The reason FDA records is "The graphical user interface may become slow in its response and in some cases non-responsive. Under some circumstances this situation may cause the ventilator to delay the annunciation of alarms". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0638-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZOLL Palm Springs
- Product
- VIASYS VELA Ventilator, Catalog Number-16186; Viasys Respiratory Care, Inc.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- The graphical user interface may become slow in its response and in some cases non-responsive. Under some circumstances this situation may cause the ventilator to delay the annunciation of alarms.
- FDA root cause
- Other
- Action
- On January 31, 2007, a product notification letter was mailed to all customers in the United States via overnight carrier (Federal Express). Consignees are informed in the notification letter that service calls are in process of being scheduled to update domestic consignees. On January 31, 2007, a product notification kit (including a product notification letter, download software, download instruction sheets and installation verification fax-back sheets) were mailed to all international customers via register mail. Consignees are informed in the notification letter that replacement software, replacement instructions, and installation verification forms are included in their recall package. Domestic Consignees will have product corrected by VIASYS service personnel. Self address postage paid envelopes will be provided with notifications to aid in effectiveness checks
- Quantity in commerce
- 878
- Distribution
- Worldwide Distribution ---USA and countries of Bangladesh, Bosnia, Brazil, China, Czech Republic, Ecuador, Estonia, Finland, France, Guatemala, Hong Kong, India, Israel, Italy, Japan, Latvia, Lithuania, Malaysia, Morocco, Netherlands, Portugal, Russian Federation, Taiwan, Thailand, Turkey & Vietnam.
- Product codes and lots
- Lot numbers: AGT03640 thru AGT06049