Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0593-2007

POOLE Suction Instrument, REF 0035040. MANUFACTURED FOR CONMED CORPORATION, 310 Broad Street, Utica, NY...

ConMed Corporation / initiated 2006-11-10 / Terminated / root cause: packaging

Conmed Corporation began a recall of POOLE Suction Instrument, REF 0035040. MANUFACTURED FOR CONMED CORPORATION, 310 Broad Street, Utica, NY 13501. ASSEMBLED IN MEXICO on . The reason FDA records is "Firm was made aware of instances where the sterile barrier of the instruments had been compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0593-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConMed Corporation
Product
POOLE Suction Instrument, REF 0035040. MANUFACTURED FOR CONMED CORPORATION, 310 Broad Street, Utica, NY 13501. ASSEMBLED IN MEXICO.
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
Firm was made aware of instances where the sterile barrier of the instruments had been compromised.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
On 11/10/06, ConMed Corp. sent "URGENT: DEVICE RECALL" letters (dated 11/10/06) to the US and foreign consignees by UPS overnight. The letters instructed the consignees to inspect their stocks of the instruments and that they return all devices. The letters also instructed the distributors to contact their customers that received the instruments and conduct sub-recalls. Additionally, when requested, ConMed notified the distributor's customers (conducted sub-recalls) by UPS overnight mail. This recall action is to the end user level. Accompanying each letter was an effectiveness check form to be completed by the consignee, and faxed back to ConMed Corp. The instruments are to be returned to ConMed Corp. in Utica, NY. For questions, the consignee can contact Ms. Nancy Crisino by phone at (315) 624-3078 or by fax at (315) 624-3089.
Quantity in commerce
2,232,896 units
Distribution
Worldwide
Product codes and lots
REF 0035040; All lots manufactured between 10/04/2001 and 10/04/2006. Lot codes on boxes and packaging contain a 7 digit lot code: For example, lot 0409271: the first 2 digits represent the year manufactured (04 represents 2004), next 2 digits represent the month (09 represents September), the next 2 digits represent the day of the month (27), and the last digit represents the manufacturing shift code (1).