Z-1003-2007
Terumo Perfusion System 8000 Roller Pump, 220/240V; Model 8000; Catalog No. 16407
Terumo Cardiovascular Systems Corporation / initiated 2004-07-01 / Terminated
Terumo Cardiovascular Systems Corp began a recall of Terumo Perfusion System 8000 Roller Pump, 220/240V; Model 8000; Catalog No. 16407 on . The reason FDA records is "Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1003-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Perfusion System 8000 Roller Pump, 220/240V; Model 8000; Catalog No. 16407.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being non-functional.
- FDA root cause
- Other
- Action
- As the complaints were received (between 07/04 and 01/06), the firm visited the customer filing the complaint and repaired the pumps demonstrating the problem. The firm will correct other units as complaints are received. The firm modified its recall strategy and issued a Device Correction Letter dated 6/22/07 notifying all users of the problem
- Quantity in commerce
- 728 pumps of all models
- Distribution
- Worldwide: USA and Bangladesh, Belgium, Bolivia, Canada, Chile, Colombia, Costa Rica, Egypt, Equador, Hong Kong, India, Indonesia, Japan, Latin America, Malaysia, Nicaragua, Philippines, Saudi Arabia, Taipe, Taiwan, United Arab Emirates, Uruguay, and Viet Nam.
- Product codes and lots
- Serial numbers 10031, 10033, 10137, 10138, 10614, 10663, 10730, 11131, 11281, 11283, 11361, 11548, 11630, 11751, 12110, 12604, 12686, 12938, 13129, 13719, 13826, 13830, 14712, 15090, 15277, 15278, 15279, 15280, 15284, 15285, 15304, 16003, 16087, 16190, 17099, 17345, 17432 and 17437 through 18122 may be affected by the problem. Serial numbers 17442, 17718, 17854, 17933, 17860, 17948, 17953, 17954, 18078, 18079 and 18107 have been corrected.