Z-0578-2007
AXIOM Artis MP, Angiographic x-ray system, model number 5904466
Siemens Medical Solutions USA, Inc. / initiated 2007-01-31 / Terminated
Siemens Medical Solutions USA, Inc began a recall of AXIOM Artis MP, Angiographic x-ray system, model number 5904466 on . The reason FDA records is "C-arm gantry could rapidly descend without command during use". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0578-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- AXIOM Artis MP, Angiographic x-ray system, model number 5904466
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- C-arm gantry could rapidly descend without command during use.
- FDA root cause
- Other
- Action
- In response to this issue, a Siemens Service Representative will visit the affected consignees beginning 02/01/2007, to check the spindle and adjust the spindle, if necessary.
- Quantity in commerce
- 62 units
- Distribution
- The products were shipped to medical facilities nationwide.
- Product codes and lots
- Model number number 5904466 - serial numbers: 20003, 20015, 20016, 20033, 20037, 20050, 20053, 20056, 20074, 20076, 20081, 20082, 20086, 20087, 20097, 20100, 20102, 20114, 20129, 20136, 20140, 20141, 20143, 20157, 20158, 20165, 20170, 20171, 20172, 20175, 20178, 20179, 20180, 20181, 20182, 20183, 20185, 20188, 20196, 20202, 20205, 20207, 20208, 20209, 20217, 20220, 20230, 20233, 20238, 20242, 20243, 20249, 20250, 20254, 20258, 20264, 20265, 20281, 20282, 20294, 20295, and 20303.