Intuitive brand Da Vinci S Surgical System, an Endoscopic Instrument Control System, Model SS2000 Patient...
Intuitive Surgical, Inc. began a recall of Intuitive brand Da Vinci S Surgical System, an Endoscopic Instrument Control System, Model SS2000 Patient Side Cart, Model PS2000, Part Number: 380267-01... on . The reason FDA records is "Da Vinci S Surgical System's spinal pin could limit mechanical motion of the arm and render the system unavailable for surgical use, and may result in system failure and the need to convert to an alternative surgical...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0670-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- Intuitive brand Da Vinci S Surgical System, an Endoscopic Instrument Control System, Model SS2000 Patient Side Cart, Model PS2000, Part Number: 380267-01, IS2000, Top Level, 4-Arm; Distributed and Manufactured by: Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- Da Vinci S Surgical System's spinal pin could limit mechanical motion of the arm and render the system unavailable for surgical use, and may result in system failure and the need to convert to an alternative surgical technique.
- FDA root cause
- Other
- Action
- The firm did not release notification to its consignees. The reworks were arranged as part of routine service visits initiated on 4/12/06.
- Quantity in commerce
- 24 units
- Distribution
- Nationwide.
- Product codes and lots
- System Numbers: SG011, SG012, SG013, SG014, SG017, SG018, SG019, SG020, SG058, and SG094.