Z-0660-2007
8mm EndoWrist Bipolar Maryland Instrument, Part Number: 400172-04, Distributed and Manufactured by Intuitive...
Intuitive Surgical, Inc. / initiated 2005-03-24 / Terminated
Intuitive Surgical, Inc. began a recall of 8mm EndoWrist Bipolar Maryland Instrument, Part Number: 400172-04, Distributed and Manufactured by Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA on . The reason FDA records is "The products were incorrectly programmed as training instruments, which allows for 30 uses (surgical procedures) instead of 10 uses". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0660-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- 8mm EndoWrist Bipolar Maryland Instrument, Part Number: 400172-04, Distributed and Manufactured by Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, CA
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- The products were incorrectly programmed as training instruments, which allows for 30 uses (surgical procedures) instead of 10 uses.
- FDA root cause
- Other
- Action
- The firm initiated its recall in March 24, 2005, when the firm sent a letter via Federal Express mail to all 18 consignees, informing them of the issue, how to identify affected product, and instructions to return the affected instruments for a full credit. The form included a reply form that consignees were requested to fill out and return.
- Quantity in commerce
- 8
- Distribution
- Product was distributed to 12 consignees in the US, who received 19 instruments, in FL, IN, LA, MN, NC, NH, NJ, NY, TN and TX. In addition, 5 consignees outside the US received product as well, in Italy, Sweden, France, Singapore and Germany.
- Product codes and lots
- The 8mm EndoWrist Bipolar Maryland Instrument is manufactured in lots, and traceability is maintained for each lot. A two digit unique identifier is added to the lot number label on the instruments, and programming records are traceable to these identifiers. The following 8 instruments from lot # 0503051 are affected: 0503051 66, 0503051 67, 0503051 68, 0503051 69, 0503051 70, 0503051 71, 0503051 72, and 0503051 73.