Z-0661-2007

MicroMaxx LAP/12-5 MHz Transducer, Laparoscopic Ultrasound Transducer

FUJIFILM SonoSite,Inc. / initiated 2007-02-05 / Terminated

Sonosite, Inc. began a recall of MicroMaxx LAP/12-5 MHz Transducer, Laparoscopic Ultrasound Transducer on . The reason FDA records is "Laparoscopic transducer exceeds the limits specified for radio-frequency emissions by up to 10db". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0661-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM SonoSite,Inc.
Product
MicroMaxx LAP/12-5 MHz Transducer, Laparoscopic Ultrasound Transducer
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
Laparoscopic transducer exceeds the limits specified for radio-frequency emissions by up to 10db.
FDA root cause
Other
Action
On 2/5/2007 firm sent by FED-Ex notification letters to firm representatives. Representatives began hand delivery of recall letter to consignees and retrieval of LAP transducer for return delivery to firm. The letter advised customers that the LAP transducer exceeded the limits specified for radio-frequency emissions. The firm also advised that exceeding the limit may cause the LAP transducer to interfere with surrounding equipment that may be susceptible to radiated emissions and cause surrounding equipment to not perform as expected.
Quantity in commerce
6 Devices
Distribution
Distribution to USA (Tennessee & Ohio), France, Korea, and Germany.
Product codes and lots
Serial Numbers: 037XT6, 0379P5, 037WMJ, 0379P1, 0379P2, 0379P0 Part Number: P05507-10