Z-0661-2007
MicroMaxx LAP/12-5 MHz Transducer, Laparoscopic Ultrasound Transducer
FUJIFILM SonoSite,Inc. / initiated 2007-02-05 / Terminated
Sonosite, Inc. began a recall of MicroMaxx LAP/12-5 MHz Transducer, Laparoscopic Ultrasound Transducer on . The reason FDA records is "Laparoscopic transducer exceeds the limits specified for radio-frequency emissions by up to 10db". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0661-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM SonoSite,Inc.
- Product
- MicroMaxx LAP/12-5 MHz Transducer, Laparoscopic Ultrasound Transducer
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Laparoscopic transducer exceeds the limits specified for radio-frequency emissions by up to 10db.
- FDA root cause
- Other
- Action
- On 2/5/2007 firm sent by FED-Ex notification letters to firm representatives. Representatives began hand delivery of recall letter to consignees and retrieval of LAP transducer for return delivery to firm. The letter advised customers that the LAP transducer exceeded the limits specified for radio-frequency emissions. The firm also advised that exceeding the limit may cause the LAP transducer to interfere with surrounding equipment that may be susceptible to radiated emissions and cause surrounding equipment to not perform as expected.
- Quantity in commerce
- 6 Devices
- Distribution
- Distribution to USA (Tennessee & Ohio), France, Korea, and Germany.
- Product codes and lots
- Serial Numbers: 037XT6, 0379P5, 037WMJ, 0379P1, 0379P2, 0379P0 Part Number: P05507-10