GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. , GE Medical...
General Electric Medical Systems Information Technology began a recall of GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. , GE Medical Systems Information Technologies, Inc., 8200 W. Tower... on . The reason FDA records is "Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0891-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Medical Systems Information Technologies, Inc.
- Product
- GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. , GE Medical Systems Information Technologies, Inc., 8200 W. Tower Ave. Milwaukee, Wisconsin USA. (System hardware consists of a processing unit and a monitor)
- Product code
- DSI Detector And Alarm, Arrhythmia
- Reason for recall
- Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when operating system memory resources reach 170MB and/or B) the system may intentionally or unexpectedly reboot .
- Root cause tag
- software
- FDA root cause
- Other
- Action
- GE Healthcare sent a certified letter titled "Urgent Medical Device Correction" to customers on 3/13/07 that explained the short and long term solutions and that the company would be providing a software update that will rectify the issue specific to CIC Pro v4.1 software.
- Quantity in commerce
- 5,190 Units
- Distribution
- Nationwide
- Product codes and lots
- All serial numbers of Software Version 4.1