Z-0733-2007

Bio-Rad-TOX/See Drug Screen Test (25 tests/box), Catalog number 194-5230; Product is distributed by Bio-Rad...

Bio-Rad Laboratories Inc. / initiated 2007-02-12 / Terminated

Bio-Rad Laboratories Inc began a recall of Bio-Rad-TOX/See Drug Screen Test (25 tests/box), Catalog number 194-5230; Product is distributed by Bio-Rad Laboratories, Inc., Hercules, CA on . The reason FDA records is "The product may produce a faint line which could be interpreted as a false negative test result". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0733-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
Bio-Rad-TOX/See Drug Screen Test (25 tests/box), Catalog number 194-5230; Product is distributed by Bio-Rad Laboratories, Inc., Hercules, CA
Product code
DKZ Enzyme Immunoassay, Amphetamine
Reason for recall
The product may produce a faint line which could be interpreted as a false negative test result .
FDA root cause
Other
Action
The firm intends to contact its subsidiary distributors and sales staff via Fax (Feb 12, 2007) and phone call. In addition, the recall notification will be posted on the firm's intranet site. Customers are instructed to destroy recalled product following standard laboratory practices.
Quantity in commerce
177 Kits
Distribution
Worldwide distribution --- USA and countries of Australia and Italy.
Product codes and lots
DOA6120272