Z-0616-2007

The iSight PACS (picture archiving and communications system) a software package used with general purpose...

Philips North America LLC / initiated 2007-02-14 / Terminated / root cause: labeling and UDI

Stentor Inc A Phillips Medical Systems Co. began a recall of The iSight PACS (picture archiving and communications system) a software package used with general purpose computer hardware to acquire, store, distribute... on . The reason FDA records is "A defect may cause patient image orientation markers to be incorrectly labeled on Multi-Planar Reformation (MPR) images. Left and Right Anatomical orientation markers may be reversed". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0616-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
The iSight PACS (picture archiving and communications system) a software package used with general purpose computer hardware to acquire, store, distribute, process, and display images/associated data throughout a clinical environment.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
A defect may cause patient image orientation markers to be incorrectly labeled on Multi-Planar Reformation (MPR) images. Left and Right Anatomical orientation markers may be reversed.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
The firm has issued a notification letter to its consignees on 02/14/2007. This notification letter contains preventive measures.
Quantity in commerce
54 units
Distribution
Worldwide Distribution ---- USA and countries of Australia, Netherlands, and Canada.
Product codes and lots
Version 3.5