The iSight PACS (picture archiving and communications system) a software package used with general purpose...
Stentor Inc A Phillips Medical Systems Co. began a recall of The iSight PACS (picture archiving and communications system) a software package used with general purpose computer hardware to acquire, store, distribute... on . The reason FDA records is "A defect may cause patient image orientation markers to be incorrectly labeled on Multi-Planar Reformation (MPR) images. Left and Right Anatomical orientation markers may be reversed". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0616-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- The iSight PACS (picture archiving and communications system) a software package used with general purpose computer hardware to acquire, store, distribute, process, and display images/associated data throughout a clinical environment.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- A defect may cause patient image orientation markers to be incorrectly labeled on Multi-Planar Reformation (MPR) images. Left and Right Anatomical orientation markers may be reversed.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- The firm has issued a notification letter to its consignees on 02/14/2007. This notification letter contains preventive measures.
- Quantity in commerce
- 54 units
- Distribution
- Worldwide Distribution ---- USA and countries of Australia, Netherlands, and Canada.
- Product codes and lots
- Version 3.5