Z-0736-2007

Roche COBAS Integra DIG, Digoxin Reagent for use on the COBAS INTEGRA models 400, 400 plus, 700 and 800...

Roche Diagnostics GmbH / initiated 2007-03-16 / Terminated / root cause: labeling and UDI

Roche Diagnostics Corp. began a recall of Roche COBAS Integra DIG, Digoxin Reagent for use on the COBAS INTEGRA models 400, 400 plus, 700 and 800 analyzers, as well as, Cobas c 501 analyzer; Catalog... on . The reason FDA records is "The lower detection limit (LDL) may be a higher value than stated in the labeling. Note: To resolve this issue, the Digoxin LDL claim has been revised to 0.3 ng/mL for all reagent lots". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0736-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche COBAS Integra DIG, Digoxin Reagent for use on the COBAS INTEGRA models 400, 400 plus, 700 and 800 analyzers, as well as, Cobas c 501 analyzer; Catalog Number 20737836322.
Product code
KXT Enzyme Immunoassay, Digoxin
Reason for recall
The lower detection limit (LDL) may be a higher value than stated in the labeling. Note: To resolve this issue, the Digoxin LDL claim has been revised to 0.3 ng/mL for all reagent lots.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
A recall letter dated 3/16/07 was sent to each consignee advising them that the LDL claim in the label has been revised to 0.3 ng/ml for all reagent lots.
Quantity in commerce
4,735
Distribution
Nationwide.
Product codes and lots
All Reagent Lots (including Lots 67941901, exp. 4/30/07; 68139201, exp. 6/30/07 and 68156201, exp. 8/31/07)