Z-0673-2007

Prostate Biopsy Needle, Model: NAC-1825BB, Soxe 18 GA x 25 CM, Single use, Ethylene sterilized...

Remington Medical / initiated 2007-01-19 / Terminated / root cause: packaging

Remington Medical Inc. began a recall of Prostate Biopsy Needle, Model: NAC-1825BB, Soxe 18 GA x 25 CM, Single use, Ethylene sterilized, Non-pyrogenic, Remington Medical, Inc., Alpharetta, GA 30005 on . The reason FDA records is "Pouches were not sealed prior to sterilization". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0673-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Remington Medical
Product
Prostate Biopsy Needle, Model: NAC-1825BB, Soxe 18 GA x 25 CM, Single use, Ethylene sterilized, Non-pyrogenic, Remington Medical, Inc., Alpharetta, GA 30005.
Product code
KNW Instrument, Biopsy
Reason for recall
Pouches were not sealed prior to sterilization.
Root cause tag
packaging
FDA root cause
Other
Action
Consignees were contacted by telephone on 01/19/2007. Each customer was informed the sterility of the product may be compromised as a result of unsealed pouches and requested to visually check their inventory.
Quantity in commerce
192 units
Distribution
Nationwide.
Product codes and lots
Lot: 063533